International Standards and Conformity Assessment for all electrical, electronic and related technologies

 

Project

Reference EN 61689:2007
Title Ultrasonics - Physiotherapy systems - Field specifications and methods of measurement in the frequency range 0,5 MHz to 5 MHz
Project Number 20792
Abstract/Scope Is applicable to ultrasonic equipment designed for physiotherapy consisting of an ultrasonic transducer generating continuous or quasi-continuous wave ultrasonic energy in the frequency range 0,5 MHz to 5 MHz. This standard only relates to ultrasonic physiotherapy equipment employing a single plane unfocused circular transducer per treatment head, producing static beams perpendicular to the face of the treatment head. This standard specifies: - methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on reference testing methods; - characteristics to be specified by manufacturers of ultrasonic physiotherapy equipment based on reference testing methods; - guidelines for safety of the ultrasonic field generated by ultrasonic physiotherapy equipment; - methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on routine testing methods; - acceptance criteria for aspects of the output of ultrasonic physiotherapy equipment based on routine testing methods. Therapeutic value and methods of use of ultrasonic physiotherapy equipment are not covered by the scope of this standard. This second edition cancels and replaces the first edition published in 1996 and constitutes a technical revision. This second edition is a result of maintenance on this standard and the referenced standards IEC 61161 (2006) and IEC 62127-1. A relatively large technical change is the determination of the effective radiating area. This is now no longer based on the measurement of four areas but only on one. This change was needed to improve the accuracy of the determination of this parameter for small transducers. Be aware that this change may alter the value obtained for this and related parameters.
Status
Published

Status

Current Stage code 6060
Current Stage code date 2007-11-07
Current Stage code deadline 2008-07-01
Deadline date for vote  

Legal

Directive(s)  
Mandate(s)  
Order Voucher  

Details

IEC Technical Body IEC/TC 87
Reference Document IEC 61689:2007 (EQV)
ICS 11.040.60 - Therapy equipment
Keywords Ultrasonics; Physiotherapy system
Note Superseded by EN 61689:2013
A-Deviation(s)  
Special National Condition(s)  

Environment

Clause
Categories - Aspects
Product Life Cycle
Information

History

Stage
Date
Deadline Date
Documents
6060 2007-11-07 2008-07-01 export to doc file  EN  DE
5099 2007-10-01 2007-12-01  
5061 2007-08-28 2007-10-01  
5060 2007-05-18 2007-06-06 export to doc file  EN
5020 2006-07-14 2006-12-15  
1090 2006-07-11 2006-07-14  

Implementation Dates

date of Ratification (DOR) (1) 2007-10-01
date of Availability (DAV) (2) 2007-11-07
date of Announcement (DOA) (3) 2008-01-01
date of Publication (DOP) (4) 2008-07-01
date of Withdrawal (DOW) (5) 2010-10-01

Relations

Supersedes EN 61689:1996
Superseded by EN 61689:2013
Normative reference (6) IEC 60050-801:1994
IEC 60469-1:1987
EN 60601-1:2006
corr. March 2010
corr.
A11:2011
EN 60601-2-5:2000
EN 61161:2007
EN 62127-1:2007
EN 62127-3:2007

(1) Date of ratification (dor) date when the Technical Board notes the approval of an EN (and HD for CENELEC), from which time the standard may be said to be approved


(2) Date of availability (dav) date when the definitive text in the official language versions of an approved CEN/CENELEC publication is distributed by the Central Secretariat


(3) Date of announcement (doa) latest date by which the existence of an EN (and HD for CENELEC), a TS or a CWA has to be announced at national level


(4) Date of publication (dop) latest date by which an EN has to be implemented at national level by publication of an identical national standard or by endorsement


(5) Date of withdrawal (dow) latest date by which national standards conflicting with an EN (and HD for CENELEC) have to be withdrawn