| Reference | EN 61689:2013 |
|---|---|
| Title | Ultrasonics - Physiotherapy systems - Field specifications and methods of measurement in the frequency range 0,5 MHz to 5 MHz |
| Project Number | 23756 |
| Abstract/Scope | IEC 61689:2013 is applicable to ultrasonic equipment designed for physiotherapy containing an ultrasonic transducer generating continuous or quasi-continuous wave ultrasound in the frequency range 0,5 MHz to 5 MHz. This standard only relates to ultrasonic physiotherapy equipment employing a single plane non-focusing circular transducer per treatment head, producing static beams perpendicular to the face of the treatment head. This standard specifies: - methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on reference testing methods; - characteristics to be specified by manufacturers of ultrasonic physiotherapy equipment based on reference testing methods; - guidelines for safety of the ultrasonic field generated by ultrasonic physiotherapy equipment; - methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on routine testing methods; - and acceptance criteria for aspects of the output of ultrasonic physiotherapy equipment based on routine testing methods. Therapeutic value and methods of use of ultrasonic physiotherapy equipment are not covered by the scope of this standard. This third edition cancels and replaces the second edition published in 2007. It constitutes a technical revision which includes the following significant technical changes with respect to the previous edition: - restriction introduced of 0,2 W/cm2 effective intensity during hydrophone measurements for treatment heads with ka≤20, to limit the likelihood of cavitation; - change in the factor Fac, to determine the effective radiating area, from 1,354 to 1,333; - and change to SI units for terms and definitions. |
| Status |
Published
|
Status
| Current Stage code | 6060 |
|---|---|
| Current Stage code date | 2013-04-19 |
| Current Stage code deadline | 2014-01-02 |
| Deadline date for vote |
Legal
| Directive(s) | |
|---|---|
| Mandate(s) | |
| Order Voucher |
Details
| IEC Technical Body | IEC/TC 87 |
|---|---|
| Reference Document | IEC 61689:2013 (EQV) |
| ICS | 11.040.60 - Therapy equipment |
| Keywords | Ultrasonics; Physiotherapy system |
| Note | Supersedes EN 61689:2007 * To be read in conjunction with EN 60601-2-5 * 2013-03-07: Publication allocated to cpalagi@cencenelec.eu |
| A-Deviation(s) | |
| Special National Condition(s) |
Environment
|
Clause
|
Categories - Aspects
|
Product Life Cycle
|
Information
|
|---|
History
Implementation Dates
| date of Ratification (DOR) (1) | 2013-04-02 |
|---|---|
| date of Availability (DAV) (2) | 2013-04-19 |
| date of Announcement (DOA) (3) | 2013-07-02 |
| date of Publication (DOP) (4) | 2014-01-02 |
| date of Withdrawal (DOW) (5) | 2016-04-02 |
Relations
| Supersedes | EN 61689:2007 |
|---|---|
| Superseded by | |
| Normative reference (6) |
EN 60601-1
IEC 60601-2-5 EN 61161:2013 EN 62127-1:2007 A1:2013 |
(1) Date of ratification (dor) date when the Technical Board notes the approval of an EN (and HD for CENELEC), from which time the standard may be said to be approved
(2) Date of availability (dav) date when the definitive text in the official language versions of an approved CEN/CENELEC publication is distributed by the Central Secretariat
(3) Date of announcement (doa) latest date by which the existence of an EN (and HD for CENELEC), a TS or a CWA has to be announced at national level
(4) Date of publication (dop) latest date by which an EN has to be implemented at national level by publication of an identical national standard or by endorsement
(5) Date of withdrawal (dow) latest date by which national standards conflicting with an EN (and HD for CENELEC) have to be withdrawn
(6) This list of normative references is purely indicative. The only official list of normative reference is the list of the published standard.